• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1100-060
Medical Device Problem Codes Off-Label Use (1494); Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a soft tissue tract for gastrostomy tube placement.The 10x60mm armada 35 balloon dilatation catheter ruptured at an unknown pressure during the first inflation.During removal resistance was felt with the device and the tip separated approximately 1cm.Fluoroscopy revealed the tip freely moving in the gastrum as well as intragastrical position of tube.It was noted the small tip fragment would pass without patient sequelae.There was no clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It was reported that the procedure was to treat a soft tissue tract for gastrostomy tube placement.It should be noted the percutaneous transluminal angioplasty catheter (pta), armada 35 / armada 35 ll, global, instruction for use (ifu) states: the device is intended for dilatation of lesions in the renal, iliac, femoral, popliteal, tibial, and peroneal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.This device is also indicated for stent post-dilatation in the peripheral vasculature.The investigation was unable to determine a conclusive cause for the reported balloon rupture; however, the reported difficulty removing the device, separation, foreign body in patient and serious injury/illness/impairment are likely due to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.D9 - device available for eval updated from 'yes' to 'no'.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14704226
Report Number2024168-2022-06444
Device Sequence Number12116270
Product Code LIT
UDI-Device Identifier08717648154959
UDI-Public08717648154959
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberB1100-060
Device Catalogue NumberB1100-060
Device Lot Number20216G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer09/06/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
Patient Weight63 KG
-
-