| Model Number |
BEA28-90/I16-30 |
| Medical Device Problem Codes |
Off-Label Use (1494); Migration (4003); Separation Problem (4043)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Rupture (2208); Stenosis (2263)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The patient was initially implanted with an afx2 bifurcated stent graft and a vela suprarenal to treat an abdominal aortic aneurysm.Approximately (4) four years post initial procedure, the patient was emitted emergently with a type 3a endoleak and rupture aorta.A secondary procedure was completed.The physician elected to implant an afx vela suprarenal.Unfortunately the following day the patient expired.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in the patient; therefore, it will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the additional endovascular procedure, ruptured aorta, type iiia endoleak of the aortic components with implant separation and death are confirmed.This is consistent with the reported adverse event/incident.User, procedure or anatomy relatedness of this adverse event could not be determined.The respiratory complications, abnormal blood loss and hemodynamic instability we all related the ruptured aorta that occurred prior to the procedure.The causation for the type iiia endoleak could not be determined.The final patient status was reported as expired on postoperative day one.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.G3: awareness date ¿ updated.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with an afx2 bifurcated stent graft and a vela suprarenal to treat an abdominal aortic aneurysm.Approximately (4) four years post initial procedure, the patient was emitted emergently with a type 3a endoleak and rupture aorta.A secondary procedure was completed.The physician elected to implant an afx vela suprarenal.Unfortunately the following day the patient expired.After the follow-up report, additional information was received and clinical evaluation also shows reasonable evidence to suggest migration of the proximal extension (12mm) and aortic stenosis (near occlusion) occurred that was not included in the event as reported.The migration and aortic stenosis were discovered during a review of the computerized tomography scan dated (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the additional endovascular procedure, ruptured aorta, type iiia endoleak of the aortic components with implant separation and death are confirmed.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest migration of the proximal extension (12mm) and aortic stenosis (near occlusion) occurred that was not included in the event as reported.The migration and aortic stenosis were discovered during a review of the computerized tomography scan dated (b)(6) 2022.The most likely cause of the type iiia endoleak with migration of the proximal stent was aortic remodeling in combination with an off-label neck.There was an increase in the aortic angulation from 30 to 83 degrees (aortic remodeling).The neck was off label as it measured at 9mm long and should be greater than or equal to 15mm.The aortic stenosis was in the severe angulation of the aorta and was related to the separation of the components.The superior stent margin at the index procedure was placed in the aortic angulation, this could have contributed to the type iiia endoleak.The respiratory complications, acute blood loss with hemodynamic instability and death were related to the ruptured aorta that occurred prior to the procedure.The final patient status was reported as expired on postoperative day one.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.B5: describe event or problem - update.G3: awareness date ¿ updated.H6: investigation type codes - remove code 4119.H6: investigation finding codes - remove code 4247.
|
| |
|