QUEST MEDICAL, INC. 10 IN. Q2 MULTIPORT MANIFOLD IV SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 9520
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received by quest medical and the complaint condition was confirmed.Quest is still pending completion of the investigation and will file a supplemental report once the investigation has been concluded.
 
Event or Problem Description
It was reported that the manifold for the multiport set was missing the top of one of the ports.It was confirmed that there were no patient complications.
 
Additional Manufacturer Narrative
Quest received the device for investigation.Analysis of the device confirmed the complaint condition reported.The root cause of the condition was inferred to be weak weld of the access port to the manifold.Quest will continue to monitor complaint trends for this complaint condition.Quest has concluded its investigation.
 
Event or Problem Description
It was reported that the manifold for the multiport set was missing the top of one of the ports.It was confirmed that there were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 IN. Q2 MULTIPORT MANIFOLD IV SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
QUEST MEDICAL, INC.
1 allentown parkway
allen TX 75002
Manufacturer (Section G)
QUEST MEDICAL, INC.
1 allentown parkway
allen TX 75002
Manufacturer Contact
brian hedden
1 allentown parkway
allen, TX 75002
9723326311
MDR Report Key14705359
Report Number1649914-2022-00012
Device Sequence Number2264856
Product Code FPA
UDI-Device Identifier00634624952003
UDI-Public00634624952003
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K151079
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2024
Device Model Number9520
Device Catalogue Number9520
Device Lot Number63133
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2022
Supplement Date Received by Manufacturer06/10/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-