ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6832
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Patient identifier complete entry = (b)(6).All available patient information was included.Additional patient details are not available.
 
Event or Problem Description
The customer observed a falsely elevated magnesium results for one patient on an architect c8000 analyzer.The following data was provided (customer¿s reference range is 1.6-2.6 mg/dl): sample id (b)(6) initial result was 3.91, repeats were 1.21 and 1.26 mg/dl.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
The complaint investigation for a falsely elevated magnesium result on an architect c8000 analyzer included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Trending review determined no related trend for the issue for the product.The historical performance of reagent lot 32185un21 was evaluated using worldwide data from abbottlink.Patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for lot 32185un21 is inside the established control limits, which indicates acceptable product performance.Device history record review did not identify any non-conformances or deviations with the likely cause lot and complaint issue.Manufacturing documentation for the likely cause lot was reviewed and did not identify any issues.Labeling was reviewed and sufficiently addresses the customer's issue.Based on the information provided and abbott diagnostics¿ complaint investigation, no systemic issue or deficiency of the magnesium reagent, lot number 32185un21, was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14705636
Report Number3002809144-2022-00209
Device Sequence Number2175764
Product Code JGJ
UDI-Device Identifier00380740161637
UDI-Public00380740161637
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/06/2022
Device Model Number03P6832
Device Catalogue Number03P68-32
Device Lot Number32185UN21
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer07/22/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date07/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/10/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C8 PROC MOD, 01G06-11, C803708.; ARC C8 PROC MOD, 01G06-11, C803708.
Patient SexUnknown
-
-