| Model Number |
SDC140H |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a mentor artoura plus smooth high profile 600cc tissue expander ruptured intraoperatively.There were no known patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned device was completed by the failure analysis lab on august 18, 2022.Device evaluation summary: a manufacturing record evaluation was performed for the finished device 9636180 number, and no non-conformance's related to the malfunction were identified.Visual analysis of the returned sample revealed that the tissue expander was found to have a tear on the anterior view measuring approximately 0.8 cm.Leak testing was performed in accordance with mentor procedures, and two (2) areas of delamination were identified at the union between the patch and shell on the posterior view.The delamination (a) was observed on the suture tab at the 5 o¿clock position causing leakage, and delamination (b) was observed on the suture tab at the 6 o' clock position with no leakage.Microscopic examination was performed on the edges of the tear, and parallel striations were found in an area of the tear, measuring approximately 0.1 cm.Parallel striations are consistent with markings made by a sharp object perforating the shell.The cause in the remaining area of the tear could not be identified.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through mentor¿s quality system.Additional investigation was performed, and it was concluded that the process controls and data records collected for two delamination (one (1) with leaks and one (1) without leaks) reported on product complaint was not able to be confirmed as manufacturing defect.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|