COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
8831661001 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during flushing, the tip of the catheter was found to be not straight (it was bent).Despite the issue observed, they still proceeded with catheter implantation and when they tried to use and send the guide wire down, there was an angle that caused the guide wire not to be delivered or advanced to the surgical site.There was nothing unusual observed on the device prior to use.There were no other products utilized with the device and the patient was treated under local anesthesia.The catheter was not repaired and there was no leak.Iodophor (iodine v) was the cleaning agent used on the device and tego was not utilized.There was no luer adapter issue and the insertion site was treated with iodophor and saline prior to product placement.The product was replaced with a new product from a different vendor to complete the surgery.There was no blood loss and blood transfusion was not required.There was no intervention/treatment required as a result of the event.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during flushing, the tip of the catheter was found to be not straight (it was bent).Despite the issue observed, the flushing was successful and they still proceeded with catheter implantation and when they tried to use and send the guide wire down, there was an angle that caused the guide wire not to be delivered or advanced to the surgical site.There was nothing unusual observed on the device prior to use.There were no other products utilized with the device and the patient was treated under local anesthesia.The catheter was not repaired and there was no leak.Iodophor (iodine v) was the cleaning agent used on the device and tego was not utilized.There was no luer adapter issue and the insertion site was treated with iodophor and saline prior to product placement.The product was replaced with a new product from a different vendor to complete the surgery.There was no blood loss and blood transfusion was not required.There was no intervention/treatment required as a result of the event.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g1, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the catheter tip was bent in a way that prevented the guidewire from being advanced.It was reported that the catheter was kinked.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|