Model Number SGC0701
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2022
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation was unable to determine a cause for the reported leak.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Event Description
This is filed to report a leak.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The steerable guide catheter (sgc) was advanced, but upon insertion into the anatomy it would not seal.Air was introduced into the device.Air did not enter the anatomy.No damage was noted.The device was removed without issue.A new sgc was used to successfully complete the procedure.One clip was implanted and mr was reduced to 1-2.There was no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14707354
MDR Text Key295485684
Report Number2024168-2022-06471
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/27/2022
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number10728R350
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/25/2022
Initial Date FDA Received06/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Age88 YR
Patient SexMale
Patient Weight72 KG
-
-