• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS AMBICOR PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS AMBICOR PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72401453
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The complaint component has not been returned; therefore, no physical or visual analysis of the product could be performed.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.Due to the lack of product return, the clinical observations of migration could not be confirmed.
 
Event or Problem Description
It was reported that the patient underwent an ambicor penile prosthesis (app) replacement surgery due to the device migration from the shaft into the body; according to the field representative, this was likely caused by patient's anatomy.A new app was implanted and no further information has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS AMBICOR PENILE PROSTHESIS
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key14707556
Report Number2124215-2022-21560
Device Sequence Number17496524
Product Code FHW
UDI-Device Identifier00878953000978
UDI-Public00878953000978
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/17/2024
Device Model Number72401453
Device Catalogue Number72401453
Device Lot Number1000373907
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2022
Initial Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/19/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age76 YR
Patient SexMale
-
-