• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL QUATTRODE LEAD WIDE SPACED, 30 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL QUATTRODE LEAD WIDE SPACED, 30 CM; SCS LEAD Back to Search Results
Model Number 3163
Medical Device Problem Codes High impedance (1291); Battery Problem: High Impedance (2947)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.During processing of this complaint, attempts were made to obtain complete patient information.
 
Event or Problem Description
Related manufacturer number: 1627487-2022-03409.It was reported the patient experienced ineffective stimulation due to high impedance.Surgery may take place later.It is unknown which lead presents the issue, therefore both leads are being reported.
 
Event or Problem Description
Additional information received indicates this lead was explanted and replaced to address the issue.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRODE LEAD WIDE SPACED, 30 CM
Common Device Name
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14709053
Report Number1627487-2022-03408
Device Sequence Number13997361
Product Code LGW
UDI-Device Identifier05414734401654
UDI-Public05414734401654
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2013
Device Model Number3163
Device Catalogue Number3163
Device Lot Number3518056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date06/15/2022
Supplement Report FDA Received Date10/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS LEAD
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight75 KG
-
-