| Model Number |
626-00-46F |
| Medical Device Problem Code |
Material Erosion (1214)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Metal Related Pathology (4530)
|
| Date of Event |
04/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
|
| |
|
Event or Problem Description
|
|
I was in a deep squat, one i had done many times before, and my implant failed and slipped out of the socket.Update per conversation with patient 05/25/2022: patient had a bilateral stryker hip replacement on (b)(6) 2018 with trident product.In (b)(6) 2022, patient was in a "deep squat" and the left hip "fell out to the back".Patient noted that prior to revision, surgeon indicated potential "stem hitting cup and liner breakage" (patient is unsure if liner was metal or poly).Patient alleges high levels of chromium and cobalt.Patient's left hip was revised on (b)(6) 2022.Patient has not returned to surgeon yet for follow up and is unsure at this time what parts were revised.Patient will supply medical records.Update 13/june/2022 wg: as reported in medwatch mw5109954: i was in a deep squat and my hip replacement failed and slipped out.I had a bilateral hip replacement with stryker trident psl on 2018.My right hip failed about 3-5 months later cause unknown, right revision in 2019.Left hip failure from a deep squat with the left hip revision surgery in 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding abnormal ion level involving an mdm liner was reported.The events were confirmed.Method & results: -product evaluation and results: the reported device was not returned however photographs were provided for review.Visual inspection of the provided photograph indicated that the device was recently explanted and covered in blood and other bodily fluids.Damage consistent with explantation and/or the disassociation event is also observed.-clinician review: a review of medical records with a clinical consultant indicated "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck." -product history review: a product history review could not be performed as no lot information was provided.-complaint history review: a complaint history review could not be performed as no lot information was provided.Conclusions: it was reported that the patient was revised due to disassociation of the head from the poly liner and abnormal ion levels.The reported device was not returned however photographs were provided for review.Visual inspection of the provided photograph indicated that the device was recently explanted and covered in blood and other bodily fluids.Damage consistent with explantation and/or the disassociation event is also observed.A review of medical records with a clinical consultant indicated "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck." no further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
I was in a deep squat, one i had done many times before, and my implant failed and slipped out of the socket.Update per conversation with patient (b)(6) 2022: patient had a bilateral stryker hip replacement on (b)(6) 2018 with trident product.In (b)(6) 2022, patient was in a "deep squat" and the left hip "fell out to the back".Patient noted that prior to revision, surgeon indicated potential "stem hitting cup and liner breakage" (patient is unsure if liner was metal or poly).Patient alleges high levels of chromium and cobalt.Patient's left hip was revised on (b)(6) 2022.Patient has not returned to surgeon yet for follow up and is unsure at this time what parts were revised.Patient will supply medical records.Update (b)(6) 2022 wg: as reported in medwatch mw5109954: i was in a deep squat and my hip replacement failed and slipped out.I had a bilateral hip replacement with stryker trident psl on 2018.My right hip failed about 3-5 months later cause unknown, right revision on 2019.Left hip failure from a deep squat with the left hip revision surgery on 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding abnormal ion level involving an mdm liner was reported.The events were confirmed.Method & results: product evaluation and results: the reported device was not returned however photographs were provided for review.Visual inspection of the provided photograph indicated that the device was recently explanted and covered in blood and other bodily fluids.Damage consistent with explantation and/or the disassociation event is also observed.Clinician review: a review of medical records with a clinical consultant indicated "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck.Trunnionosis can also cause elevated ions" product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the poly liner and abnormal ion levels.The reported device was not returned however photographs were provided for review.Visual inspection of the provided photograph indicated that the device was recently explanted and covered in blood and other bodily fluids.Damage consistent with explantation and/or the disassociation event is also observed.A review of medical records with a clinical consultant indicated "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck.Trunnionosis can also cause elevated ions" no further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
I was in a deep squat, one i had done many times before, and my implant failed and slipped out of the socket.Update per conversation with patient (b)(6) 2022: patient had a bilateral stryker hip replacement on (b)(6) 2018 with trident product.In (b)(6) 2022, patient was in a "deep squat" and the left hip "fell out to the back".Patient noted that prior to revision, surgeon indicated potential "stem hitting cup and liner breakage" (patient is unsure if liner was metal or poly).Patient alleges high levels of chromium and cobalt.Patient's left hip was revised on (b)(6) 2022.Patient has not returned to surgeon yet for follow up and is unsure at this time what parts were revised.Patient will supply medical records.Update 13/june/2022 wg: as reported in medwatch mw5109954: i was in a deep squat and my hip replacement failed and slipped out.I had a bilateral hip replacement with stryker trident psl on 2018.My right hip failed about 3-5 months later cause unknown, right revision on 2019.Left hip failure from a deep squat with the left hip revision surgery on 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding abnormal ion level involving an mdm liner was reported.The events were confirmed.Method & results: -product evaluation and results: the reported device was not returned however photographs were provided for review.Visual inspection of the provided photograph indicated that the device was recently explanted and covered in blood and other bodily fluids.Damage consistent with explantation and/or the disassociation event is also observed.Clinician review: a review of medical records with a clinical consultant indicated "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck.Trunnionosis can also cause elevated ions" -product history review: review of the device history records indicate devices were manufactured and accepted into final stock on 12/21/2017 with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the poly liner and abnormal ion levels.The reported device was not returned however photographs were provided for review.Visual inspection of the provided photograph indicated that the device was recently explanted and covered in blood and other bodily fluids.Damage consistent with explantation and/or the disassociation event is also observed.A review of medical records with a clinical consultant indicated "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck.Trunnionosis can also cause elevated ions" the patient inquired on any recalls associated with these devices and a search indicated there are no recalls associated with these parts/lots.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
I was in a deep squat, one i had done many times before, and my implant failed and slipped out of the socket.Update per conversation with patient (b)(6) 2022: patient had a bilateral stryker hip replacement on (b)(6) 2018 with trident product.In (b)(6) 2022, patient was in a "deep squat" and the left hip "fell out to the back".Patient noted that prior to revision, surgeon indicated potential "stem hitting cup and liner breakage" (patient is unsure if liner was metal or poly).Patient alleges high levels of chromium and cobalt.Patient's left hip was revised on (b)(6) 2022.Patient has not returned to surgeon yet for follow up and is unsure at this time what parts were revised.Patient will supply medical records.Update 13/june/2022 wg: as reported in medwatch mw5109954: i was in a deep squat and my hip replacement failed and slipped out.I had a bilateral hip replacement with stryker trident psl on 2018.My right hip failed about 3-5 months later cause unknown, right revision on 2019.Left hip failure from a deep squat with the left hip revision surgery on 2022.
|
| |
|