| Model Number |
DA5115ST |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was returned to the manufacturer for analysis.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
It was reported that when negative pressure was applied to a sofia catheter, air bubbles appeared near a pinhole leak neat the hub.It was not used and there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned device was not able to replicate bubbles, however further examination of the device found that there was a gap in adhesive which was clogged with dried blood.Air could've flowed through the hole to created bubbles at the time of the reported event.This findings from this investigation have been communicated to quality.
|
| |
|