• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG SL-PLUS/SBG/IPA MIA OFFSET ADAPTER 10MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG SL-PLUS/SBG/IPA MIA OFFSET ADAPTER 10MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Model Number 75007309
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a thr surgery, a piece of the spring device that allows the attachment of sl-plus/sbg/ipa mia offset adapter 10mm broke off.The procedure was completed, with a non-significant delay, using a s+n back-up device.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
It was reported that during a thr surgery, a piece of the spring device that allows the attachment of sl-plus/sbg/ipa mia offset adapter 10mm broke off.The device in question was returned for investigation.Upon visual inspection, the reported failure mode regarding the broken-off spring could be confirmed.The missing spring was not returned.A complaint history review was performed.No other complaints regarding the batch in question were found.The review of the production documentation did not detect any deviations that could have contributed to the reported failure mode.The performance of the device is within the risks, which are anticipated in the risk management documentation of the product, where the corresponding failure mode and the severity are covered.A review of past corrective actions was performed.No further escalation is required.According to the performed investigations, the relationship between the reported event and the device was confirmed.No indications that the reported device failed to meet manufacturing specifications upon release for distribution were found.The root cause for the fracture could not be identified and stays undetermined.The need for corrective action is not indicated.We remind to follow our document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19) that all devices must be inspected and controlled for proper functioning after cleaning/disinfection.Nevertheless, smith + nephew will continue to monitor the device for similar issues.The returned device will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL-PLUS/SBG/IPA MIA OFFSET ADAPTER 10MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14711062
Report Number9613369-2022-00303
Device Sequence Number17496541
Product Code KWY
UDI-Device Identifier07611996076394
UDI-Public07611996076394
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75007309
Device Catalogue Number75007309
Device Lot NumberG54925
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer09/08/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date09/09/2022
Date Device Manufactured05/14/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-