Model Number 254625
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer in china that during an acl (ant.Cruciate ligament) reconstruction procedure on (b)(6) 2022, it was observed that the bio-intrafix tapered screw, 8-10 mm x 30 mm anchor device had a crack on it.Another like device was used to complete the surgery.There were no adverse consequences to the patient nor surgical delay reported.No additional information could be provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available fo the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation, however a photo was provided.Upon visual inspection of the photo, the photo shows the bio-intrafix cracked at the middle of the tapered screw.Also, it could be observed biological residues on the anchor.Hands on analysis should provide the evidence necessary to confirm the root cause.A manufacturing record evaluation was performed for the finished device lot number: 8l45054, and no nonconformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint can be confirmed.The broken tapered screw can be attributed to procedural variables, such handling of the device or product interaction during procedure.The possible root cause could be related to a failure to engage until fully seated position of the tapered screw during implantation.However, it cannot be conclusively affirmed.As per ifu, should insert hex driver into tapered screw to a fully seated position.Failure to engage screw completely may result in damage to tapered screw during implantation.Do not insert a 6¿8 mm screw into a tunnel less than 8 mm in diameter.Use the hard bone sizing with tibialis allografts.Failure to insert the screw along the tunnel axis may result in device damage.Guidewire use is recommended.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received and evaluated.Visual inspections revealed marks of use on the device.It could be observed the bio-intrafix was broken into two pieces at the middle of the tapered screw.Also, it could be observed biological residues on the anchor.A manufacturing record evaluation was performed for the finished device lot number: 8l45054, and no nonconformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.The broken tapered screw can be attributed to procedural variables, such handling of the device or product interaction during procedure.The possible root cause could be related to a failure to engage until fully seated position of the tapered screw during implantation.However, it cannot be conclusively affirmed.As per ifu, should insert hex driver into tapered screw to a fully seated position.Failure to engage screw completely may result in damage to tapered screw during implantation.Do not insert a 6¿8 mm screw into a tunnel less than 8 mm in diameter.Use the hard bone sizing with tibialis allografts.Failure to insert the screw along the tunnel axis may result in device damage.Guidewire use is recommended.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-INTRAFIX TAPERED SCREW, 8-10 MM X 30 MM
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14711210
Report Number1221934-2022-01767
Device Sequence Number11313180
Product Code HWC
UDI-Device Identifier10886705013227
UDI-Public10886705013227
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K032167
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number254625
Device Catalogue Number254625
Device Lot Number8L45054
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer06/24/2022
07/21/2022
07/26/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date06/27/2022
07/22/2022
07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-