• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT CURRENT PLUS VR, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT CURRENT PLUS VR, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1211-36Q
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During an in-clinic follow-up, a rapid drop in battery depletion was noted on the device.Premature battery depletion was suspected.The device was explanted and replaced to resolve the event.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported field event of premature battery depletion was confirmed by analysis.The device was at elective replacement indicator (eri) when received.A longevity calculation was performed and found the battery depletion was premature based on the device usage.Electrical testing confirmed high current drain from the hybrid.A hybrid anomaly was found to be the cause of the high current drain and premature depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT PLUS VR, DF-4 CONNECTOR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14712382
Report Number2017865-2022-11685
Device Sequence Number17496427
Product Code NVZ
UDI-Device Identifier05414734503716
UDI-Public05414734503716
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2017
Device Model NumberCD1211-36Q
Device Catalogue NumberCD1211-36Q
Device Lot NumberS000009790
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date10/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/04/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-