Model Number 1521-90-059
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The primary surgery was performed on (b)(6) 2022.The revision surgery with the total hip arthroplasty was performed on (b)(6) 2022.During the revision surgery, when the surgeon removed the inner head, black ring was found on both mating parts of the stem and the head.Uneven wear and oxidation discoloration was found on the liner sliding surfaces.After removal of the cup, the zb¿s advantage was used for the acetabulum side.Although the sales rep explained regarding the risks of using delta ceramic head, the surgeon firmly requested to use the delta ceramic head.Therefore, the delta ceramic head was used as inner head under the surgeon¿s responsibility.Doi: (b)(6) 2022; dor: (b)(6) 2022; unknown hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9/10 COCR HEAD 28MM +3
Common Device Name
SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14712737
Report Number1818910-2022-10999
Device Sequence Number5233424
Product Code JDI
UDI-Device Identifier10603295276647
UDI-Public10603295276647
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1521-90-059
Device Catalogue Number152190059
Device Lot NumberWVM40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer06/29/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date07/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
9/10 COCR HEAD 28MM +3.; AML A PLUS 12.0 MM.; DURALOC 1200 SERIES 50MM OD.; ENDURON 10D 50OD X 28ID.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-