DEPUY ORTHOPAEDICS INC US ARTICULEZE M HEAD 36MM +12; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-54-000
Medical Device Problem Codes Degraded (1153); Naturally Worn (2988)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Inflammation (1932); Necrosis (1971); Ambulation or Postural Difficulties (2544); Metal Related Pathology (4530); Muscle/Tendon Damage (4532); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the patient "i had a hip replacement in 2014 that failed earlier this year.I am trying to find out if the parts that were used were recalled or should have not been used in combination.The part was depuy summit with a competitor titanium cup and cobalt chrome on polyethylene articulation." doi: 2014 : dor: none reported.Affected side: unknown.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: b5, h6 (clinical code).H6 clinical code: code for blood heavy metal increased was updated to metal related pathology (e1618).
 
Event or Problem Description
Primary surgery occurred on 10/01/2014.Sticker sheets for primary implants are located within the sticker sheet attachment.The femoral side is depuy and the acetabular side is competitor.Revision operative notes (b)(6) 2020 indicated that the patient received a right total hip revision due to unspecified loosening, alval soft tissue reaction with chronic inflammation and necrotic tissue.And implant corrosion.It was noted that the femoral head was worn with corrosion between the head junction of the stem.The surgery was completed without indication of complication by the surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULEZE M HEAD 36MM +12
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14713453
Report Number1818910-2022-11017
Device Sequence Number13997242
Product Code JDI
UDI-Device Identifier10603295033974
UDI-Public10603295033974
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K980513
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Model Number1365-54-000
Device Catalogue Number136554000
Device Lot Number7926369
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/30/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +12; COMPETITOR CUP (STRYKER); COMPETITOR LINER(STRYKER); SUMMIT POR TAPER SZ6 STD OFF
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-