Model Number 16-02-85 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The patient pump and te distribution board were replaced and the problem could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message associated to a patient pump during procedure.There was no patient injury.
|
|
Manufacturer Narrative
|
H.10: the most likely root cause of the reported event is a pump circuit board defect.This led to the damage on the safety resistors of the distribution board.
|
|
Event Description
|
See initial report.
|
|