• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. POLYAXIAL SCREW; SIZE Ø4.5X40 MM; POSTERIOR CERVICAL SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. POLYAXIAL SCREW; SIZE Ø4.5X40 MM; POSTERIOR CERVICAL SCREW SYSTEM Back to Search Results
Model Number 7601-04540
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Status and location of the device is unknown.
 
Event or Problem Description
It was reported that two yukon oct spinal system polyaxial screws heads broke off intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the first of the two screws.
 
Event or Problem Description
It was reported that two yukon oct spinal system polyaxial screws heads broke off intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.This report captures the first of the two screws.
 
Additional Manufacturer Narrative
Device analysis could not be performed as the device was not returned.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.A review of complaint history associated with the subject catalog was performed, no adverse trends were observed.Limited information regarding the event was made available to stryker.Multiple attempts were made to obtain additional information, but no response was received.A root cause cannot be established with the information available.If additional information is received or the device is returned for evaluation, the investigation will be reopened and updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL SCREW; SIZE Ø4.5X40 MM
Common Device Name
POSTERIOR CERVICAL SCREW SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14713776
Report Number3004774118-2022-00240
Device Sequence Number14019465
Product Code NKG
UDI-Device Identifier10888857349865
UDI-Public10888857349865
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K171444
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7601-04540
Device Catalogue Number7601-04540
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight91 KG
-
-