| Model Number |
PROV45 |
| Medical Device Problem Code |
Misassembled (1398)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
| Date of Event |
05/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2022 during a left atrial appendage (laa) exclusion procedure, t was noticed that the dacron was not completely covering the prov45 clip.An injury to the laa was observed.The patient was placed back on-pump and the procedure was converted to a sternotomy to repair the laa.An atriclip pro1 was placed on the the laa successfully.There were no additional patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) the device was not returned for investigation however, the product history record was reviewed for prov45 lot number: 94661.There is nothing in the device history record that would indicate that the devices were released with any non-conformances that would contribute to the complaint.
|
| |
|