• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RADIAL JAW 4; FORCEPS, BIOPSY, NON-ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RADIAL JAW 4; FORCEPS, BIOPSY, NON-ELECTRIC Back to Search Results
Model Number M00513401
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a radial jaw 4 standard capacity biopsy forceps was used in a procedure performed on (b)(6)2022.During preparation, a hair was found on the device.The device was not used and another radial jaw 4 biopsy forceps was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Block h6: device code a180104 captures the reportable event of foreign material present in device.Block h10: the returned radial jaw 4 standard capacity biopsy forceps pouch was analyzed, and a visual evaluation observed that the pouch was completely unopen.A foreign material that looks like a hair was found inside the pouch.As per the tappi chart, the foreign material is greater than tolerated.The reported event was confirmed.Based on all available information, the problem was traced to manufacturing process.Therefore, the investigation conclusion code is manufacturing deficiency as the foreign material inside the pouch corresponds to a deficiency in the process.An investigation to address this issue is in progress.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a radial jaw 4 standard capacity biopsy forceps was used in a procedure performed on (b)(6) 2022.During preparation, a hair was found on the device.The device was not used and another radial jaw 4 biopsy forceps was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIAL JAW 4
Common Device Name
FORCEPS, BIOPSY, NON-ELECTRIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14717978
Report Number3005099803-2022-03216
Device Sequence Number16694904
Product Code FCL
UDI-Device Identifier08714729777960
UDI-Public08714729777960
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/16/2025
Device Model NumberM00513401
Device Catalogue Number1340
Device Lot Number0029060188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-