Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested, but unavailable: were x-rays performed to localize the needle? no further information is available.Was the needle removed from the patient? no further information is available.Procedure name? no further information is available.Device return status? the device has been received at sukagawa and will be shipped.Please check rmao.No further information will be provided.
|
|
Product complaint # (b)(4).Date sent to the fda: 7/11/2022.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: one empty open foil, a needle and suture product code y443 was returned to ethicon for analysis.Upon visual analysis of the returned sample revealed that light swage defect was observed on the swage area.The barrel hole was examined under 20x magnification and no remnant suture was found.In addition, the suture end was examined and there isn¿t sufficient impression, resulting in a needle pull off defect.Due to returned condition of the sample cannot functional test was performed.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|