(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained via: "please clarify, was the suture broken and the needle separated from the suture during the procedure? yes.Was the suture broken or frayed during the procedure? no.Did the event occur during one or multiple patient procedures? one procedure.What is the total number of procedures? one.If this event occurred in multiple procedures, please create a product complaint for each event and provide the respective reference number(s).Sample analysis: the product was returned for evaluation.Visual inspection was conducted on the returned sample.Visual analysis of the returned sample determined that an empty labeled winding former and a needle-suture of product code 8559h were received for evaluation.An incorrect swage was noted since the marks of the swage area was misaligned (one stake near the suture) and consequently, the suture is severely cut.Based on the information currently available, the incorrect swage was identified during the investigation of the sample received.This product issue will be addressed through ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
|