SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 22003E-07
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd alaris smartsite extension set experienced occlusion.The following information was provided by the initial reporter: extension set will not flush.
 
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
Investigation summary: one sample model 22003e-07 was returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The set was attached to a cut off bd syringe to see if a fluid path can be observed while the piston is in the activated position.A fluid path was observed while the piston was in the activated position.The smartsite was then connected to a 10 ml bd syringe filled with blue dye water and the set was attempted to be flushed.The set was successfully flushed.The customer complaint that the extension set will not flush could not be replicated.The root cause could not be determined because the issue could not be replicated.A device history record review for model 22003e - 07 lot number 22029678 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 11feb2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A trend for this occlusion issue has been identified for this product line.A capa (corrective action preventative action) has been initiated, and a team has been assembled in order to investigate the issue.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Capa# : pr # (b)(4).
 
Event or Problem Description
It was reported that the bd alaris smartsite extension set experienced occlusion.The following information was provided by the initial reporter: extension set will not flush.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14721572
Report Number9616066-2022-00722
Device Sequence Number5195908
Product Code FPA
UDI-Device Identifier50885403242043
UDI-Public50885403242043
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/10/2023
Device Model Number22003E-07
Device Catalogue Number22003E-07
Device Lot Number22029678
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer06/21/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-