• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GREENLIGHT XPS LASER SYSTEM; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GREENLIGHT XPS LASER SYSTEM; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-0210R
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that error 302.13 and 832 were received on the console.The procedure was not completed due to the event and the patient had been sedated with general anesthesia at the time.There were no patient complications.This event is being reported for aborted/cancelled procedure with a patient under anesthesia.
 
Additional Manufacturer Narrative
The console was not returned for analysis.The console was inspected onsite by the field service engineer (fse) was able to confirm and duplicate errors 302.13 and 832.While troubleshooting, he was able to determine that the errors were the result of an issue with the connecting cable between the top cover and master control board (mcb).The top cover was replaced, and the console passed full functional and electrical safety testing.The top cover was then returned.Upon receipt of this top cover at our quality assurance laboratory, this device was thoroughly analyzed.Visual analysis of the top cover found major scuff marks on the enclosure from normal usage wear.Functional testing of the top cover confirmed error 832 interlock error code was received, indicating an electrical fault with the touch screen board.Based on analysis results, the most probable cause of the reported event was an electrical circuit fault.The reported event is confirmed.
 
Event or Problem Description
It was reported that error 302.13 and 832 were received on the console.The procedure was not completed due to the event and the patient had been sedated with general anesthesia at the time.There were no patient complications.This event is being reported for aborted/cancelled procedure with a patient under anesthesia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT XPS LASER SYSTEM
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key14723856
Report Number2124215-2022-21733
Device Sequence Number14022341
Product Code GEX
UDI-Device Identifier00878953001906
UDI-Public00878953001906
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K092735
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0010-0210R
Device Catalogue Number0010-0210R
Device Lot NumberXPS50510
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/09/2022
Supplement Date Received by Manufacturer11/01/2022
Initial Report FDA Received Date06/16/2022
Supplement Report FDA Received Date11/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/13/2012
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-