| Model Number |
DFR00V |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Flashers (1864)
|
| Date of Event |
10/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the preloaded intraocular lens (iol) was explanted from patient's right eye as the patient was seeing starbursts at night while driving.Another lens of different model and lower diopter was used as the replacement.There was no patient injury and suture was placed.The patient was happy with the vision post-operation.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient weight: unknown, information not provided.Device evaluated by mfr: the device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: 11/20/2024.Section h3: device evaluated by manufacturer: yes.Device evaluation: product evaluation was not performed because the suspect product has not been received.If product is received after initial closure, the pi and cp (if necessary) may be re-opened to complete the return investigation.The complaint issue reported could not be confirmed.Conclusion: based on the complaint investigation results, the product was released within specifications.The complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.Based on the complaint investigation results, the product was released within specifications.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|