• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 TITANIUM LG SUB-Q-PORT K; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 TITANIUM LG SUB-Q-PORT K; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 0602230
Medical Device Problem Codes Fracture (1260); Stretched (1601); Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiration date: 06/2025).
 
Event or Problem Description
It was reported that during a port placement procedure, the guidewire of the port allegedly had resistance and failure to advance.It was further reported that the guidewire was twisted.The procedure was completed using another device.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a port placement procedure, the guidewire of the port allegedly had resistance and failure to advance.It was further reported that the guidewire was twisted.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation, one electronic photo was provided for review.The investigation is confirmed for the reported guidewire stretched issue as the guidewire was noted to be raveled & uncoiled on the proximal end near to the guidewire tip.However, the investigation is inconclusive for the reported failure to advance and physical resistance issue as the exact circumstances at the time of the reported event was unknown and cannot be verified.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 06/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a port placement procedure, the guidewire of the port allegedly had resistance.It was further reported that the guidewire was twisted.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.One electronic photo was provided for review.The investigation is confirmed for the reported guidewire stretched ad identified core wire fracture issues as the guidewire was noted to be raveled & uncoiled on the proximal end near to the guidewire tip.Furthermore, upon zooming the provided photo, one of the core wire was appeared to be fractured.However, the investigation is inconclusive for the reported physical resistance issue as the exact circumstances at the time of the reported event was unknown and cannot be verified from the provided photo.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiration date: 06/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITANIUM LG SUB-Q-PORT K
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14726957
Report Number3006260740-2022-02305
Device Sequence Number20229598
Product Code LJT
UDI-Device Identifier00801741025563
UDI-Public(01)00801741025563
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K050310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0602230
Device Catalogue Number0602230
Device Lot NumberREFQ0217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer07/30/2022
12/07/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date08/03/2022
12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-