• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE 2.4C 8.0T 1050L

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE 2.4C 8.0T 1050L Back to Search Results
Model Number EG-2490K
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Air/water nozzle missing.
 
Additional Manufacturer Narrative
We checked the returned unit and confirmed that the air/water nozzle missing.Based on the result, we concluded that it was caused due to the physical damage applied on the air/water nozzle.In addition, we confirmed that the operation channel perforated, the air/water nozzle clogged, the air/water nozzle missing, and the operation channel leaky; however, they are not the main cause, and/or irrelevant to the alleged complaint.As a result of confirming the detail as gfe, no response was obtained, so we can not deny the possibility of the nozzle dropped into human body.Therefore, based on the technical report ""hr-rpt-0585(nozzle)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
VIDEO GASTROSCOPE 2.4C 8.0T 1050L
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key14728433
Report Number9610877-2022-54691
Device Sequence Number6047974
Product Code FDS
UDI-Device Identifier04961333152858
UDI-Public04961333152858
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131902
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEG-2490K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2022
Initial Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-