| Model Number |
ETLW1624C93E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section information references the main component of the system.Other relevant device(s) are: product id: etlw1620c124e, serial/lot #: (b)(4), ubd: 06-sep-2019, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii stent graft system was implanted during an unknown endovascular procedure.It was noted that the patient has high vessel tortuosity and mild vessel calcification both at implant and currently.It was reported the patient returned for a follow-up ct approximately 4 years and 8 months after the index procedure where thrombus was observed inside both left and right iliac limbs.The physician relined the left limb using a 16-16-82 limb and right limb with a 16-24-124 limb.As per the physician the cause of the stent thrombus is unknown. no additional clinical sequelae was provided and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received: it was confirmed there was no limb occlusion, just the presence of thrombus.The physician wanted to reline the limbs to trap the thrombus.It was confirmed the thrombus was contained in the limbs only and did not extend proximally.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|