The customer reported to olympus, a routine microbiological culture was performed on the uretero-reno fiberscope.The scope tested positive for one (1) colony forming unit (cfu) of filamentous fungi.There was no reported contamination or any other serious deterioration in the state of health of any person, to which the scope could have been a contributory cause.
|
The cleaning, disinfection, and sterilization (cds) of the scope was performed by the customer.The air/water channel, auxiliary channel, balloon channel, and forceps elevator wire channel were flushed.There was an aspiration of water through the instrument/suction channel as well.The instrument/suction channel, suction cylinder, instrument channel port, balloon channel, and distal end/areas around the elevator were manually cleaned with hexanios g + r pre-disinfectant detergent.The scope was manually disinfected with anioxyde 1000.The scope was stored horizontally.The maintenance of the scope was performed by olympus.After the device was returned to olympus, it was sent out for additional testing.The hygiene microbiological investigation report indicated the channels, and the distal end of the scope were cultured.Less than one (1) cfu of a microorganism was detected.The results obtained complied with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.The subject device has been received and is currently in the evaluation process.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was evaluated and no abnormalities were found.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 4 years since the subject device was manufactured.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with instructions for use (ifu), the results conformed to the regulation's recommendation.The following is included in the instructions for use (ifu): "warning: an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|