| Catalog Number |
14-589607 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Two polaris 4.75 dominos were returned with the pieces reported on 3012447612-2022-00086 through 3012447612-2022-00095, and were also found to have damaged threads.There were no reported patient impacts.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3012447612-2022-00157.
|
| |
|
Event or Problem Description
|
|
Two polaris 4.75 dominos were returned with the pieces reported on 3012447612-2022-00086 through 3012447612-2022-00095, and were also found to have damaged threads.There were no reported patient impacts.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection revealed damage to the threading of all the returned devices.There was no damage to the screw drives.Root cause: root cause was unable to be determined.This event could possibly be attributed to inserting the closure tops off-axis and cross threading them, and then continuing to attempt installation on cross-threaded screws.Dhr review: per dhr review, the parts were likely conforming when they left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|