Catalog Number 300865
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/17/2022
Event Type  malfunction  
Event Description
It was reported while using bd plastipak¿ syringes leakage occurred from a hole.There was no report of patient impact.The following information was provided by the initial reporter: date of occurrence: (b)(6) 2022.Incident description: a department reported to us a leak of norepinephrine solution during syringe preparation.On visual inspection clearly shows that the body of the syringe was already damaged and that this syringe should have been immediately discarded.
 
Manufacturer Narrative
No photos or physical samples that display the reported condition were available for investigation.A device history review was performed for the reported lot 2201089, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures, no issues related to the reported incident were identified during manufacturing.Damage can occur due to the product jamming within manufacturing equipment.Without the sample to evaluate, we cannot verify this to be true for the incident reported.Based on the available information we cannot identify a root cause at this time.Complaints received for this device and reported condition will be monitored by our quality team for signs of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PLASTIPAK¿ SYRINGES
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14733675
MDR Text Key294531361
Report Number3003152976-2022-00274
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number300865
Device Lot Number2201089
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/20/2022
Initial Date FDA Received06/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-