| Model Number |
45-652050 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformance associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was received and the investigation ongoing.Once the investigation is completed, a supplement report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a splenic artery aneurysm embolization procedure, the coil had resistance when it was advanced through the glide catheter.When the physician pulled the pusher wire back to remove it, the coil detached.The glide catheter and coil were removed together successfully.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned coil system found the implant stretched at the proximal end with deformed loops and separated from the pusher.No indication using a detachment controller was found on the pusher's heater coil.The investigation found the pusher's monofilament with a tensile break shape at the tip which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
|
| |
|