| Catalog Number |
302833 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reported phone #: (b)(6).Investigation summary: a device history record review was completed by our quality engineer team for provided material number 302833 and lot number 8214901.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that there was leakage past the stopper/plunger while using bd 30 ml luer slip syringe.This occurred 8 times, however, no patient impact reported.The following information was provided by the initial reporter, translated from chinese to english: the nurse in charge found fluid leakage at the connection between the syringe and the piston when drawing the normal saline.
|
| |
|
Search Alerts/Recalls
|