ABBOTT VASCULAR HI-TORQUE SPARTACORE GUIDE WIRE Back to Search Results
Model Number 1005203
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional armada 14 device is filed under separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat the left peroneal artery with right groin access.A spartacore guide wire was advanced and the 2.0x200mm armada 14 balloon catheter was advanced.There was no difficulty removing the stylet or protective sheath and the balloon was prepared/flushed prior to use.The balloon was attempted to be inflated but there was a lot of resistance.It took a lot of pressure to inflate the balloon and it was only partially inflated and not uniformly shaped.The device was losing pressure but there was no leak of contrast noted in the vessel as the balloon was inflating, nor was there any blood backfilling the inflation device so it was difficult to tell if there was a rupture.When attempting to remove the armada balloon, it became stuck with the spartacore guide wire in the superficial femoral artery and they were pulled out together, causing wire access to be lost.There was no attempt made to move the guide wire while the balloon was pressurized/inflated.The physician noted a thread of what seemed to be plastic material at the hub of the armada sheath and pulled it out without intervention.It was a long plastic material but angiography post procedure confirmed nothing remained in the patient.A command18 guide wire was advanced and a 1.5x40mm armada 14 and 2.5x200mm armada balloon catheters were used to complete the procedure.Angiogram followed revealing in line flow down to foot.There was no adverse patient effect, however, there was a delay in the procedure but no harm to the patient.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported difficult to remove was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that the noted damages to the armada 14 balloon catheter (inner member tear) resulted in the guide wire poking through the inner member tear resulting in the reported difficult to remove.Manipulation of the device resulted in the noted core folded back.The noted multiple core bends likely occurred due to handling or during packing for return analysis.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE SPARTACORE GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14735236
Report Number2024168-2022-06561
Device Sequence Number12273992
Product Code DQX
UDI-Device Identifier08717648014031
UDI-Public08717648014031
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K963702
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model Number1005203
Device Catalogue Number1005203
Device Lot Number2021462
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/27/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARMADA 14 BALLOON CATHETER
Patient Age67 YR
Patient SexMale
Patient Weight67 KG
-
-