• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US HEALIX KNOTLESS AWL, 3.4MM; BONE AWL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US HEALIX KNOTLESS AWL, 3.4MM; BONE AWL Back to Search Results
Model Number 219495
Medical Device Problem Codes Device Contamination with Body Fluid (2317); Illegible Information (4050)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer in australia that during an unknown surgery on (b)(6) 2022, it was observed that the depth markings on the healix knotless awl, 3.4mm have worn off.During in-house engineering evaluation of the photo provided by the customer, it was determined that the depth laser etch mark was faded and it could be observed biological residues (blood) on the device.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation, however a photo was provided.Upon visual inspection of the photo, the photo shows a healix knotless awl 3.4mm with the depth laser etch mark faded and it could be observed biological residues (blood) on the device.Hands on analysis should provide the evidence necessary to confirm the root cause.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint can be confirmed.Since this is a reusable device, the constant manipulation between surgeries and sterilization process can lead to a laser etch mark to faded.However, it cannot be conclusively affirmed.As per ifu, clean instruments as soon as possible after use.If cleaning must be delayed, wet the instruments in a compatible liquid solution to prevent drying and encrustation of surgical soil.Avoid prolonged exposure to saline to minimize the chance of corrosion.Remove excessive soil with a disposable wipe.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection was found use wear on the device.It could be observed laser etch mark faded.Finally, no structural anomalies were found on the device.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.Since this is a reusable device, the constant manipulation between surgeries and sterilization process can lead to a laser etch mark to faded.However, it cannot be conclusively affirmed.As per instructions for use (ifu), clean instruments as soon as possible after use.If cleaning must be delayed, wet the instruments in a compatible liquid solution to prevent drying and encrustation of surgical soil.Avoid prolonged exposure to saline to minimize the chance of corrosion.Remove excessive soil with a disposable wipe.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX KNOTLESS AWL, 3.4MM
Common Device Name
BONE AWL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14735303
Report Number1221934-2022-01840
Device Sequence Number19439748
Product Code LXH
UDI-Device Identifier10886705022373
UDI-Public10886705022373
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number219495
Device Catalogue Number219495
Device Lot Number1306001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer07/12/2022
07/18/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date07/13/2022
07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-