| Model Number |
CREAJ |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable crej2 creatinine jaffé gen.2 results for two samples from one patient.The customer questioned if the patient's medications/treatments could have affected the results.Sample 1 was drawn at 01:56 pm and the results were 6.16 mg/dl, 6.25 mg/dl, and 6.59 mg/dl.Sample 2 was drawn at 5:12 pm and the results were 0.27 mg/dl and 0.31 mg/dl.The questionable results were not reported outside of the laboratory.The customer used cobas 6000 c501 module serial number (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The analyzer alarm trace contained several abnormal probe sucking alarms on the date of the event.This is usually an indicator of a pre-analytical handling issue.The provided data for calibration and quality control were within the specification.Therefore a general reagent and/or system-related issue could be excluded.The investigation did not identify a product problem.The cause of the event could not be determined as no sample material was available for further investigation.
|
| |
|