Model Number A20976A
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
The olympus representative reported on behalf of the customer, during a procedure, the bridge was swiveling during a case.After the procedure, at reprocessing, it was noted the bridge unscrewed apart.The two pieces of the bridge were dismantled and bioburden, possibly crusted blood, was found in the bridge.The facility took apart the other six (6) bridges and bioburden was found on the other six (6) bridges as well.No patient harm reported.This complaint is related to patient identifiers: (b)(6).
 
Additional Manufacturer Narrative
Model #: 167 - 0103 or 162 - 0103 or 16z - 0103.Lot number was partially illegible.In speaking with olympus technical support via the phone, no troubleshooting was performed.The olympus representative stated the customer requested feedback from olympus on how to move forward, now that this bioburden was found inside the devices.The facility had owned the sets for multiple years and never disassembled them as the process was not indicated in the instructions for use.Since the issue had occurred, the facility had removed the bioburden from each bridge and were taking the bridges apart for each procedure to ensure the bridges were clean.The subject device referenced in this report was not returned for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Additional Manufacturer Narrative
This is a supplemental report to correct the initial mdr (9610773 - 2022 - 00224).Upon further review, it was determined this complaint/report is a duplicate of mfr report number # 9610773 - 2022 - 00223.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRIDGE, ONE WAY
Common Device Name
BRIDGE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key14736213
Report Number9610773-2022-00224
Device Sequence Number2226868
Product Code GCP
UDI-Device Identifier04042761020398
UDI-Public04042761020398
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA20976A
Device Catalogue NumberA20976A
Device Lot Number167 - 0103 OR 162 OR 16Z
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer06/20/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-