| Model Number |
M00576560 |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on may 27, 2022 that an ultraflex tracheobronchial covered distal release stent was to be implanted to treat a malignant intrinsic tumor in the left main stem bronchus during a stent placement procedure performed on (b)(6) 2022.The patient's anatomy was tortuous and was dilated prior to stent placement.During the procedure, the physician applied a lot of force to deploy the ultraflex stent.The stent was fully deployed but, a portion of the stent was lodged in the endotracheal (et) tube.The et tube along with the stent were removed completely.The patient was re-intubated and the procedure was completed with another ultraflex tracheobronchial stent.There were no patient complications reported as a result of this event.
|
| |
|