Model Number 1456Q/75
Medical Device Problem Codes Intermittent Capture (1080); Capturing Problem (2891)
Health Effect - Clinical Codes Discomfort (2330); Syncope/Fainting (4411)
Date of Event 05/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information was requested, but not received.
 
Event or Problem Description
Related manufacturer reference number: 2017865-2022-12804; related manufacturer reference number: 2017865-2022-12805.It was reported that a patient presented to the emergency room reporting falls and syncopial episodes.Upon examination, the patient¿s right ventricular (rv) and left ventricular (lv) lead were found to have intermittent loss of capture.The loss of capture was alleged on the patient¿s pacemaker.The rv and lv leads were found to also have high capture thresholds.Corrective programming changes were made.The patient was stable throughout.
 
Event or Problem Description
New information received notes that intermittent capture was alleged on the right ventricular (rv) and left ventricular (lv) leads.The event occurred on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 75 CM
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14738268
Report Number2017865-2022-12807
Device Sequence Number19389838
Product Code NIK
UDI-Device Identifier05414734510141
UDI-Public05414734510141
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number1456Q/75
Device Catalogue Number1456Q/75
Device Lot NumberA000116341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age71 YR
Patient SexFemale
Patient Weight70 KG
-
-