• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. INTERTAN 10S 10MM X 18CM 125D; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. INTERTAN 10S 10MM X 18CM 125D; NAIL, FIXATION, BONE Back to Search Results
Model Number 71675383
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Necrosis (1971)
Date of Event 06/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after an internal fixation surgery with an intertan short nail had been performed on 2018, the patient experienced avascular necrosis of femoral head.The hardware (short nail, lag/compression screw and 5.0 locking screw) was removed on (b)(6) 2022 prepare for a total hip arthroplasty.Patient is recovering.
 
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device shows signs of prolonged use.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.The clinical/medical evaluation concluded that based on the information provided and the product evaluation the reported avascular necrosis could not be confirmed.Although age-related wear could not be ruled out a likely contributory factor.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could not be corroborated.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction or loss of sterility.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.Corrected information: h6 (health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10S 10MM X 18CM 125D
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14738567
Report Number1020279-2022-03042
Device Sequence Number5583124
Product Code JDS
UDI-Device Identifier00885556105696
UDI-Public00885556105696
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K040212
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/17/2025
Device Model Number71675383
Device Catalogue Number71675383
Device Lot Number15LM15340
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient SexFemale
-
-