| Model Number |
4054 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported: "faulty, intermittent light".No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: a sensor was returned without the "do not discard" tag and evaluated.The sensor passed eeprom and visual inspection.During continuity testing, an open in cable on the emitter circuit at the bend relief area of sensor end was found.The sensor failed functional testing which resulted in the lab monitor to display an error message when connected with the sensor.The sensor led did not illuminate.
|
| |
|
Event or Problem Description
|
|
The customer reported: "faulty, intermittent light".No consequences or impact to patient were reported.
|
| |
|