SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP TRIAL INSERT NAV 22/53; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 75023361
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, a polarcup trial insert nav 22/53 broke while impacting.Surgery was finished with a sn back-up device and no delay was reported.No harm to the patient or further complications reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: it was reported that a polarcup trial insert nav 22/53 broke while impacting.Surgery was finished with a sn back-up device and no delay was reported.No harm to the patient or further complications reported.The complaint device, used in treatment, was returned for investigation and the reported failure mode could be confirmed upon visual inspection.The complaint device was found fractured.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.A review of the batch record revealed no deviations from the standard manufacturing process.A review of the complaint history revealed no additional complaint for the batch in question.A review of the risk management documentation verifies the failure mode and severity of the reported issue.No prior escalation actions are related to the reported issue.Based on the performed investigations, the reported failure mode could be confirmed.A relationship between the reported event and the device can be confirmed.The reported device met manufacturing specifications upon release for distribution.No probable cause can be determined.Normal wear and tear through repeated use is known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.This version of the device will be monitored for similar issues.This investigation is considered closed.The complaint device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP TRIAL INSERT NAV 22/53
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14739009
Report Number9613369-2022-00311
Device Sequence Number16697990
Product Code LXH
UDI-Device Identifier07611996112504
UDI-Public07611996112504
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K110135
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75023361
Device Catalogue Number75023361
Device Lot NumberA52845
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/05/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-