• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH 200, LLC TOWEL OR BLUE STERILE; FIBER, MEDICAL, ABSORBENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH 200, LLC TOWEL OR BLUE STERILE; FIBER, MEDICAL, ABSORBENT Back to Search Results
Model Number 28700-006
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
Customer reported the or towels were disintegrating after getting wet in the sterile saline in the sterile bowls, during a watchman left atrial appendage occlusion device procedure.Fuzzy type pieces were coming off.There was no harm to the patient.
 
Additional Manufacturer Narrative
Based on supplier investigation, device history record (dhr) review did not indicate any exception that could lead to the reported incident.The average linting data is 0.105g/10pcs.A sample was not returned for investigation.According to supplier, the or towel is made of cotton, so cotton fiber is born.Supplier is continuously working with cardinal health to better control the linting and have implemented several measures to improve it: suction machines have been installed in grey cloth rolling process, dyeing process and cutting process, the suction process was added before product's final folding, and workers do it according to standard operation procedure requirement, linting test method and acceptable criteria was stipulated to see the suction results.(=0.38g/10pieces), and in the folding process, supplier used one cloth pad under 100 pieces semi-finished products to avoid linting stuck onto the products during product's transfer.From the investigation, there is no abnormal situation happened in production or dhr.Therefore, the root cause could not be determined.Cardinal will continue to monitor trends for this product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOWEL OR BLUE STERILE
Common Device Name
FIBER, MEDICAL, ABSORBENT
Manufacturer (Section D)
CARDINAL HEALTH 200, LLC
3651 birchwood dr
waukegan IL 60085
Manufacturer (Section G)
CARDINAL HEALTH SHANGHAI
f20, tower 1, kerry everbright
shanghai 20007 0
CH   200070
Manufacturer Contact
patricia tucker
3651 birchwood drive
waukegan, IL 60085
8478874151
MDR Report Key14739144
Report Number1423537-2022-00792
Device Sequence Number12174617
Product Code FRL
UDI-Device Identifier10630140917257
UDI-Public10630140917257
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number28700-006
Device Catalogue Number28700-006
Device Lot Number220106-18-SH
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/24/2022
Event Location Hospital
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/24/2022
Initial Report FDA Received Date06/17/2022
Date Device Manufactured01/24/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-