| Model Number |
TRM-240 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
04/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture baker grade iii is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture, capsular contracture baker grade iii, and exchange from textured to smooth.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture, capsular contracture baker grade iii, and exchange from textured to smooth breast implants due to the patient¿s concern with the product.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: underweight, red particles in the shell and opening on posterior.A visual and microscopic analysis was performed which identified: sharp opening on posterior.A dimension measurement in the shell was performed which identify the thickness within specification.Base on the device analysis the final assessment is: sharp opening on posterior assessed as an unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture, capsular contracture baker grade iii, and exchange from textured to smooth breast implants due to the patient¿s concern with the product.Device has been explanted.
|
| |
|