• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/24/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pt info: unknown/ asked but not available.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.The device has not been returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Health effect - clinical code (b)(4).Attempts have been made to obtain the missing information.However, to date it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Event or Problem Description
Customer reported having a patient issue.They indicated sack torn while inserting an intraocular lens (iol) into a patient's left eye.The iol was cut out and replaced.The incision had to be enlarged.It is unclear whether the iol caused the capsule tear.The replacement lens is a different johnson & johnson iol model za9003, +19.5 diopter.No further information was provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: jul 8, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received stuck to the original lens case and cut in half.The lens was removed from the case and cleaned, revealing that the lens was also torn.No further issues could be observed with the lens.Based on the return condition of the lens, no further product evaluation could be performed.A product quality deficiency could not be determined.The complaint issue "capsule tear", "removal and replacements", and "incision enlargement" could not be identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14739186
Report Number3012236936-2022-01615
Device Sequence Number12196423
Product Code HQL
UDI-Device Identifier05050474531703
UDI-Public(01)05050474531703(17)260316
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/16/2026
Device Model NumberZCB00
Device Catalogue NumberZCB0000195
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer09/18/2022
Initial Report FDA Received Date06/17/2022
Supplement Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE 1MTEC30, LOT CK09123
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-