| Model Number |
ORT200 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported they observed hydraulic fluid on the operating room table covers.No patient involvement or malfunction in the table's operation was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An imris customer service engineer went on-site to investigate the reported leakage of hydraulic fluid.The engineer observed pooling of hydraulic fluid inside the table covers and traced the cause of the leak to an o-ring between the table's hydraulic pump housing and fluid reservoir.No patient involvement or malfunction of the table's hydraulic system was reported.The engineer replaced the o-ring and verified table function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer in response to an internally identified issue.
|
| |
|