• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number S-55-100-120-P6
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that during stent deployment interaction with the anatomy and/or other devices resulted in the ratchet being unable to properly engage/fully deploy the stent; thus resulting in reported difficult or delayed activation; however this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was performed to treat a target lesion in the proximal superficial femoral artery.The supera stent delivery system was advanced to the target lesion.During deployment of the supera stent, the ratchet mechanism did not push the stent forward; it did not seem to be engaging appropriately.The physician made continued attempts, and the supera stent was finally able to be deployed.It was noted that the stent was implanted more proximal than was desired, and extended into the profunda artery.The target lesion was fully covered.Post dilatation was performed per standard protocol.No additional intervention was performed.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14739457
Report Number2024168-2022-06592
Device Sequence Number17592534
Product Code NIP
UDI-Device Identifier08717648226120
UDI-Public08717648226120
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model NumberS-55-100-120-P6
Device Catalogue NumberS-55-100-120-P6
Device Lot Number0092361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/27/2022
Initial Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
Patient Weight86 KG
-
-