Model Number 74022224
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a tka surgery when the box of the journey tibia base np lt sz 4 was opened, the sterile packaging was damaged, it looked like a sterilization or over heating issue.The outer packaging looked fine.The procedure was completed, with a non-significant surgical delay, using a s+n back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.This issue was evaluated through our internal quality process and determined there is no objective evidence that this was a manufacturing error that happened here at smith-nephew, therefore no additional evaluation should be initiated.The sterile seal barrier was not broken and part was not exposed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Damage during shipping or mishandling process errors could be probable causes of the reported event.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIA BASE NP LT SZ 4
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14739792
Report Number1020279-2022-03052
Device Sequence Number14964403
Product Code JWH
UDI-Device Identifier03596010554321
UDI-Public03596010554321
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74022224
Device Catalogue Number74022224
Device Lot Number22CM09058
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date06/18/2022
Supplement Report FDA Received Date08/03/2022
Date Device Manufactured03/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-