• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIDEX CORPORATION G-PROBE; LASER INSTRUMENT, SURGICAL, POWERED, PRODUCT CODE: HQF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIDEX CORPORATION G-PROBE; LASER INSTRUMENT, SURGICAL, POWERED, PRODUCT CODE: HQF Back to Search Results
Model Number 15980
Health Effect - Clinical Code Eye Burn (2523)
Date of Event 05/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Iridex became aware of a complaint reporting a patient experiencing multiple conjunctival burns during treatment with the g-probe.A specific root cause could not be determined.Based on similar reported events, the most likely cause of the issue is suspected to be contamination (e.G.Blood, tissue or betadine) on the probe tip when the laser was activated or laser activation when the probe is over pigmented conjunctival tissue.The is no additional information available at this time.Engineering failure analysis was not performed because the device was not returned for evaluation.The g-probe ifu has clear instructions to examine the probe tip before treatment and to avoid areas of perilimbal pigmentation.Per the ifu 'excessive treatment power may result in ocular surface burns or ciliary body hemorrhage.Contamination of the fiber optic tip by blood or tissue char may result in ocular surface burns.Excessive energy may cause equatorial burns.Heavy perilimbal conjunctival pigmentation may result in local absorption and burns; therefore, avoid areas of heavy perilimbal pigmentation.' if the devices are returned and the results of failure analysis indicate new information, the adverse event report will be amended.
 
Event or Problem Description
Iridex became aware of a complaint reporting a patient experiencing multiple conjunctival burns during treatment with the g-probe.A specific root cause could not be determined.Based on similar reported events, the most likely cause of the issue is suspected to be contamination (e.G.Blood, tissue or betadine) on the probe tip when the laser was activated or laser activation when the probe is over pigmented conjunctival tissue.The is no additional information available at this time.The g-probe ifu has clear instructions to examine the probe tip before treatment and to avoid areas of perilimbal pigmentation.Per the ifu 'excessive treatment power may result in ocular surface burns or ciliary body hemorrhage.Contamination of the fiber optic tip by blood or tissue char may result in ocular surface burns.Excessive energy may cause equatorial burns.Heavy perilimbal conjunctival pigmentation may result in local absorption and burns; therefore, avoid areas of heavy perilimbal pigmentation.".
 
Additional Manufacturer Narrative
This follow-up report is provided to update the following: 1)part b: b5 -product problem section was updated.2) report c: is this a combination product - no.3) report d: updates to section d8, d9, d4.4) part h: update to include that this a correction and additional information report 5) part d: updated this section with the udi number.6) part e: updates to the initial reporter section with information regarding the complainant.Note: patient information was not provided, therefore part a is not filled out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G-PROBE
Common Device Name
LASER INSTRUMENT, SURGICAL, POWERED, PRODUCT CODE: HQF
Manufacturer (Section D)
IRIDEX CORPORATION
1212 terra bella ave.
mountain view CA 94043 1824
Manufacturer (Section G)
IRIDEX CORPORATION
1212 terra bella ave
mountain view CA 94043 9149
Manufacturer Contact
madhumita srikanth
1212 terra bella ave
mountain view, CA 94043-9149
6502189149
MDR Report Key14740054
Report Number2939653-2022-00019
Device Sequence Number14663519
Product Code HQF
UDI-Device Identifier00813125015350
UDI-Public813125015350
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K143154
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number15980
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/18/2022
Supplement Report FDA Received Date05/03/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-