EXACTECH, INC. ACUMATCH GXL 15DEG LINER 36MM SZ H; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number ACUMATCH GXL 15DEG LINER 36MM SZ H
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately twelve years post initial left tha, the 66 y/o male patient had a revision due to poly worn over time.Patient was revised with new acumatch 36mm size h 15 degree liner.Head also replaced with biolox delta 36mm +0.Pt deemed stable.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Photos received.The device is not available due to hospital policy.
 
Additional Manufacturer Narrative
Concomitant medical products: cocr fem head 36mm +0 offset 12/14 (cat# 142-36-00 / serial# (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
The revision reported was likely the result of a combination of long-term prosthesis wear and patient conditions secondary to being implanted for 12 years.However, this cannot be confirmed as the device was not available for evaluation.The most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUMATCH GXL 15DEG LINER 36MM SZ H
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14740092
Report Number1038671-2022-00686
Device Sequence Number11331712
Product Code JDI
UDI-Device Identifier10885862023391
UDI-Public10885862023391
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K051556
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/04/2012
Device Model NumberACUMATCH GXL 15DEG LINER 36MM SZ H
Device Catalogue Number132-36-28
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer07/29/2022
Initial Report FDA Received Date06/19/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/11/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
-
-